Experience of using a non-invasive pulse-wave transit time-based cardiac output monitoring in patients undergoing robot-assisted surgery with pneumoperitoneum in a head-down position

https://doi.org/10.15022/00004779
https://doi.org/10.15022/00004779
9d7cfeb6-c2b6-4bf7-87a3-f15bf448a56c
名前 / ファイル ライセンス アクション
kiyoum-38-1-38to42.pdf kiyoum-38-1-38to42 (771.4 kB)
Item type 紀要論文 / Departmental Bulletin Paper(1)
公開日 2020-02-18
タイトル
タイトル Experience of using a non-invasive pulse-wave transit time-based cardiac output monitoring in patients undergoing robot-assisted surgery with pneumoperitoneum in a head-down position
言語
言語 eng
キーワード
言語 en
主題Scheme Other
主題 continuous cardiac output, head-down position, pnuemoperitoneum
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ departmental bulletin paper
ID登録
ID登録 10.15022/00004779
ID登録タイプ JaLC
著者 Kurota, Misato

× Kurota, Misato

Kurota, Misato

Search repository
Takahashi, Tatsuro

× Takahashi, Tatsuro

Takahashi, Tatsuro

Search repository
Akimoto, Ryo

× Akimoto, Ryo

Akimoto, Ryo

Search repository
Nakane, Masaki

× Nakane, Masaki

Nakane, Masaki

Search repository
Kawamae, Kaneyuki

× Kawamae, Kaneyuki

Kawamae, Kaneyuki

Search repository
著者別名
識別子Scheme WEKO
識別子 14656
姓名 黒田, 美聡
著者別名
識別子Scheme WEKO
識別子 14657
姓名 高橋, 達朗
著者別名
識別子Scheme WEKO
識別子 14658
姓名 秋元 , 亮
著者別名
識別子Scheme WEKO
識別子 14659
姓名 中根, 正樹
著者別名
識別子Scheme WEKO
識別子 14660
姓名 川前, 金幸
内容記述
内容記述タイプ Other
内容記述 Background: Estimated continuous cardiac output(esCCO)is a non-invasive technique for monitoring continuous cardiac output(CCO)that is based on pulse-wave transit time. In this study, we report the use of CCO monitoring for patients undergoing robot-assisted surgery with pneumoperitoneum in a head-down position.
Method: Thirteen patients undergoing radical robot-assisted prostatectomy under general anesthesia were enrolled. Intraoperatively, esCCO and arterial pressure-based cardiac output(APCO)were simultaneously recorded. The association between esCCO and APCO was then evaluated using correlation analysis and Bland-Altman analysis. The trending ability of esCCO was evaluated by 4-quadrant plot analysis.
Results: The correlation coefficient between esCCO and APCO was 0.54(P < 0.0001). According to the Bland-Altman plot, the bias and precision values were 1.38 and 0.79 L/min, respectively. The concordance analysis showed the concordance rate of 92.3%.
Conclusion: These results indicate that esCCO is capable of tracking hemodynamic changes associated with pneumoperitoneum in the head-down position.
書誌情報 山形大学紀要. 医学 : 山形医学
en : Bulletin of the Yamagata University. Medical science : Yamagata medical journal

巻 38, 号 1, p. 38-42, 発行日 2020-02-15
ISSN
収録物識別子タイプ ISSN
収録物識別子 0288-030X
書誌レコードID
収録物識別子タイプ NCID
収録物識別子 AN00067279
その他の言語のタイトル
その他のタイトル 頭低位、気腹を伴うロボット支援手術における、脈波伝播時間を用いた非侵襲的心拍出量計の使用経験
タイトル(ヨミ)
その他のタイトル トウテイイ キフク ヲ トモナウ ロボット シエン シュジュツ ニ オケル ミャクハ デンパン ジカン ヲ モチイタ ヒシンシュウテキ シンハクシュツリョウケイ ノ シヨウ ケイケン
出版者
出版者 山形大学
出版者(ヨミ)
値 ヤマガタ ダイガク
別言語の出版者
値 Yamagata University
資源タイプ
内容記述タイプ Other
内容記述 論文(Article)
資源タイプ・ローカル
値 紀要論文
資源タイプ・NII
値 Departmental Bulletin Paper
資源タイプ・DCMI
値 text
資源タイプ・ローカル表示コード
値 2
戻る
0
views
See details
Views