Experience of using a non-invasive pulse-wave transit time-based cardiac output monitoring in patients undergoing robot-assisted surgery with pneumoperitoneum in a head-down position
https://doi.org/10.15022/00004779
https://doi.org/10.15022/000047799d7cfeb6-c2b6-4bf7-87a3-f15bf448a56c
名前 / ファイル | ライセンス | アクション |
---|---|---|
kiyoum-38-1-38to42 (771.4 kB)
|
|
Item type | 紀要論文 / Departmental Bulletin Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2020-02-18 | |||||
タイトル | ||||||
タイトル | Experience of using a non-invasive pulse-wave transit time-based cardiac output monitoring in patients undergoing robot-assisted surgery with pneumoperitoneum in a head-down position | |||||
言語 | ||||||
言語 | eng | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | continuous cardiac output, head-down position, pnuemoperitoneum | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | departmental bulletin paper | |||||
ID登録 | ||||||
ID登録 | 10.15022/00004779 | |||||
ID登録タイプ | JaLC | |||||
著者 |
Kurota, Misato
× Kurota, Misato× Takahashi, Tatsuro× Akimoto, Ryo× Nakane, Masaki× Kawamae, Kaneyuki |
|||||
著者別名 | ||||||
識別子Scheme | WEKO | |||||
識別子 | 14656 | |||||
姓名 | 黒田, 美聡 | |||||
著者別名 | ||||||
識別子Scheme | WEKO | |||||
識別子 | 14657 | |||||
姓名 | 高橋, 達朗 | |||||
著者別名 | ||||||
識別子Scheme | WEKO | |||||
識別子 | 14658 | |||||
姓名 | 秋元 , 亮 | |||||
著者別名 | ||||||
識別子Scheme | WEKO | |||||
識別子 | 14659 | |||||
姓名 | 中根, 正樹 | |||||
著者別名 | ||||||
識別子Scheme | WEKO | |||||
識別子 | 14660 | |||||
姓名 | 川前, 金幸 | |||||
内容記述 | ||||||
内容記述タイプ | Other | |||||
内容記述 | Background: Estimated continuous cardiac output(esCCO)is a non-invasive technique for monitoring continuous cardiac output(CCO)that is based on pulse-wave transit time. In this study, we report the use of CCO monitoring for patients undergoing robot-assisted surgery with pneumoperitoneum in a head-down position. Method: Thirteen patients undergoing radical robot-assisted prostatectomy under general anesthesia were enrolled. Intraoperatively, esCCO and arterial pressure-based cardiac output(APCO)were simultaneously recorded. The association between esCCO and APCO was then evaluated using correlation analysis and Bland-Altman analysis. The trending ability of esCCO was evaluated by 4-quadrant plot analysis. Results: The correlation coefficient between esCCO and APCO was 0.54(P < 0.0001). According to the Bland-Altman plot, the bias and precision values were 1.38 and 0.79 L/min, respectively. The concordance analysis showed the concordance rate of 92.3%. Conclusion: These results indicate that esCCO is capable of tracking hemodynamic changes associated with pneumoperitoneum in the head-down position. |
|||||
書誌情報 |
山形大学紀要. 医学 : 山形医学 en : Bulletin of the Yamagata University. Medical science : Yamagata medical journal 巻 38, 号 1, p. 38-42, 発行日 2020-02-15 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0288-030X | |||||
書誌レコードID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AN00067279 | |||||
その他の言語のタイトル | ||||||
その他のタイトル | 頭低位、気腹を伴うロボット支援手術における、脈波伝播時間を用いた非侵襲的心拍出量計の使用経験 | |||||
タイトル(ヨミ) | ||||||
その他のタイトル | トウテイイ キフク ヲ トモナウ ロボット シエン シュジュツ ニ オケル ミャクハ デンパン ジカン ヲ モチイタ ヒシンシュウテキ シンハクシュツリョウケイ ノ シヨウ ケイケン | |||||
出版者 | ||||||
出版者 | 山形大学 | |||||
出版者(ヨミ) | ||||||
値 | ヤマガタ ダイガク | |||||
別言語の出版者 | ||||||
値 | Yamagata University | |||||
資源タイプ | ||||||
内容記述タイプ | Other | |||||
内容記述 | 論文(Article) | |||||
資源タイプ・ローカル | ||||||
値 | 紀要論文 | |||||
資源タイプ・NII | ||||||
値 | Departmental Bulletin Paper | |||||
資源タイプ・DCMI | ||||||
値 | text | |||||
資源タイプ・ローカル表示コード | ||||||
値 | 2 |